DQR ID | Subject | Affected Data Streams | Data Quality Metric |
---|---|---|---|
D170708.2 | SGP/CO2FLX/C1 - Failed thermopile sensor | sgpco2flxrad4mC1.b1 | Incorrect |
Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
2017-05-01 | 08:23:00 | 2017-06-29 | 16:15:00 | Incorrect |
Subject: | SGP/CO2FLX/C1 - Failed thermopile sensor |
---|---|
Affected Datastreams: | sgpco2flxrad4mC1.b1 |
Description | It looks like one of the seven miniature thermopile sensors in the SPN1 went bad. The sensor was swapped with an AmeriFlux spare SPN1 sensor on 6/29 by site technician. The faulty sensor was sent to the manufacturer for repair under. |
Measurements |
|