DQRID : D040819.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/02/1995183009/03/19950130Suspect
08/14/1995183008/15/19950130Suspect
09/13/1995183009/14/19950130Suspect
08/31/1995183009/01/19950130Suspect
09/03/1995183009/04/19950130Suspect
08/29/1995183008/30/19950130Suspect
08/15/1995183008/16/19950130Suspect
08/30/1995183008/31/19950130Suspect
08/28/1995183008/29/19950130Suspect
09/04/1995183009/05/19950130Suspect
08/20/1995183008/21/19950130Suspect
more
Subject:
SGP/MWR/B1 - Loss of thermal stabilization
DataStreams:sgpmwrlosB1.a1, sgpmwrlosB1.b1, sgpmwrB1.00, sgpmwrlosB1.00, sgp5mwravgB1.c1,
sgpmwrlosB1.a0
Description:
Periodically during August and September 1995 the microwave radiometer generated error 
messages in the Site Operations Log. This error indicates that the temperature of the 
microwave hardware (specifically, the cross-coupler or "xc") exceeded its controlled 
temperature (47-52 degC) at which point it was no longer thermally stabilized and the gain was 
uncontrolled.  During these periods which typically last about 8 hours from about local noon 
until about sunset the data behave anomalously and should be considered invalid.

Specifically, the precipitable water vapor increases and the liquid water path decreases 
(and becomes SIGNIFICANTLY NEGATIVE (-0.1 mm) on clear sky days).  The RMS noise level in 
the data also increases sharply.  The 'Tkxc' field appears ONLY in the a0-level data and 
does NOT appear in either the a1 (mwrlos) or c1 (mwr5avg) files.  Therefore THESE 
ANOMALOUS VALUES HAVE BEEN INCLUDED IN THE 5-MINUTE AVERAGES.

The microwave hardware is thermally stabilized to about +/- 0.1 degC by resistive heating. 
 When the internal temperature rises above the set point the thermal stabilization of 
the instrument gain is lost. From an examination of the component temperature data, it 
appears that increasing the set point temperature to about 55 degC (328 K) would prevent a 
re-occurrence of this problem at the SGP.  The manufacturer, Radiometrics concurs that 
raising the set point will fix this problem and will not cause other problems.
Suggestions: 
Do not use data
Measurements:sgpmwrB1.00:
  • Raw data stream - documentation not supported
more
sgp5mwravgB1.c1:
  • 31tbsky
  • 23tbsky
  • vap
  • liq
more
sgpmwrlosB1.a1:
  • 31tbsky
  • 23tbsky
  • vap
  • liq
more
sgpmwrlosB1.00:
  • Raw data stream - documentation not supported
more
sgpmwrlosB1.a0:
  • 31tbsky
  • 23tbsky
  • liq
  • vap
more
sgpmwrlosB1.b1:
  • tbsky31
  • liq
  • tbsky23
  • vap
more

Close this window