Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/18/1995 | 0130 | 04/18/1995 | 1620 | Incorrect |
Subject: | SGP/EBBR/E9 - Stopped Ventilation Fan |
DataStreams: | sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1 |
Description: | On April 18, 1995, a site operator discovered the right aspirator ventilation fan of the EBBR to be stopped. Inspection of the data indicates that this condition began early in the GMT day. Interpretation of the start time of the condition is made more difficult by a slight bias low in absolute temperature in comparison to the left temperature probe. Lack of ventilation of the temperature and relative humidity sensors attached to the AEM results in incorrect measurements of air temperature and relative humidity by the right aspirator probes, and therefore all quantities calculated from them, including sensible and latent heat fluxes, are incorrect as well. The quantities that are incorrect for the period above are: 5 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, and vp_bot 15 minute: rr_thum_r, mv_hum_r, tair_r 30 minute: tair_top, tair_bot, thum_top, thum_bot, hum_top, hum_bot, vp_top, vp_bot, bowen, e, h |
Suggestions: | |
Measurements: | sgp30ebbrE9.b1: more
sgp15ebbrE9.b1: more sgp5ebbrE9.b1: more |