DQRID : D950626.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/18/1995113004/18/19951530Incorrect
more
Subject:
SGP/EBBR/E9 - Water in Net Radiometer
DataStreams:sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1
Description:
The top dome of the net radiometer was broken and a small amount of water
had collected in the bottom dome of the net radiometer.  It is not obvious
that it had been there for more than several hours before the site operators
arrived for a preventative maintenance visit.  There are noticable effects on
the data for the period listed above.  The domes were replaced.  The following
data is incorrect for the period above:

5 minute:  q

15 minute: mv_q

30 minute: q, e, h
Suggestions: 
Measurements:sgp30ebbrE9.b1:
  • latent_heat_flux
  • net_radiation
  • sensible_heat_flux
more
sgp15ebbrE9.b1:
  • net_radiation_mv
more
sgp5ebbrE9.b1:
  • net_radiation
more

Close this window