DQRID : D960104.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/19/1995104012/19/19951650Incorrect
more
Subject:
SGP/EBBR/E7 - Wind Speed Cups Frozen
DataStreams:sgp15ebbrE7.b1, sgp5ebbrE7.b1, sgp30ebbrE7.b1
Description:
Freezing rain, snow, and below freezing conditions during 19 January
1995 caused the wind speed sensor cups to freeze in a stopped state at E7,
Elk Falls, KS on 19 January 1995.  When temperatures rose to above freezing
later that day, the ice thawed and the cups turned freely again.

Measurements affected by the stalled wind speed sensor include:

5 minute:  wind_s, res_ws

15 minute: wind_s

30 minute: wind_s, res_ws
Suggestions: 
Measurements:sgp5ebbrE7.b1:
  • wspd_arith_mean
  • wspd_vec_mean
more
sgp15ebbrE7.b1:
  • wspd_arith_mean
more
sgp30ebbrE7.b1:
  • wspd_vec_mean
  • wspd_arith_mean
more

Close this window