DQRID : D961122.8
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/24/1996171509/24/19961730Incorrect
more
Subject:
SGP/EBBR/E12 - Left Air Temperature Thermocouple Incorrect
DataStreams:sgp5ebbrE12.b1, sgp30ebbrE12.b1, sgp15ebbrE12.b1
Description:
The left air temperature thermocouple was offscale, for unknown reasons, for 
1715 GMT for 15 minute, and 1730 GMT for 30 minute air temperature values.

The malfunction of the air temperature thermocouple results in incorrect Bowen 
ratio, sensible heat flux, and latent heat flux during the period indicated. 

The data values that are affected are:

5 minute:  tair_top, tair_bot

15 minute: tair_l

30 minute: tair_top, tair_bot, bowen, e, and h
Suggestions: 
Measurements:sgp15ebbrE12.b1:
  • temp_air_left
more
sgp5ebbrE12.b1:
  • temp_air_top
  • temp_air_bottom
more
sgp30ebbrE12.b1:
  • temp_air_top
  • sensible_heat_flux
  • latent_heat_flux
  • temp_air_bottom
  • bowen_ratio
more

Close this window