Subject:
SGP/MWR/C1 - IRT Calibration
DataStreams:sgpmwrlosC1.b1, sgpmwrlosC1.a1
Description:
Comparison of the uplooking IRT with AERI for 5 May 1997 during a time
when the 9-11 micrometer sky temperature was in the range 220-228 K,
according to AERI, the IRT indicated approximately 3 K higher.  This is
outside of the specified uncertainty in the IRT and indicates that the
calibration needs to be revised.
Suggestions: 
The instrument calibration must be checked.