DQRID : D980527.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/16/1997201006/20/19971313Incorrect
more
Subject:
SGP/AOS/C1  - Single-wavelength nephelometer errors
DataStreams:sgpaosC1.a0
Description:
Due to a malfunctioning bypass valve, the single wavelength nephelometer was not properly 
sampling ambient air.  Thus, scattering values during this period are incorrect.  The 
incorrect data field is "Bscat530nm".
Suggestions: 
Measurements:sgpaosC1.a0:
  • Bscat530nm
more

Close this window