Subject:
SGP/SIRS/E2 - Reprocess: wrong calibration
DataStreams:sgpsirsE2.01, sgpsirsE2.a1
Description:
Mike Splitt found the following in the Site Ops OMIS data base from E2 on
11/26/97 :


        Found that the wrong calibration sticker was placed on the nip
        and the wrong Cal factor had been entered in the program. Also
        while editing the program I found that I had the wrong Cal
        factor for the DS. I inserted the correct Cal factors for the DS
        and Nip, complied and down loaded the new program into
        SIRS_02 at 1720GMT on 11/26/97

This entry was made by Craig Webb.

The problem was reported in P971121.1
Suggestions: 
These data need to be reprocessed.