Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/22/2000 | 2330 | 11/24/2000 | 1400 | Incorrect |
Subject: | NSA/MWR/C1 - Reprocess: calibration change |
DataStreams: | nsamwrlosC1.b1, nsamwrlosC1.a1 |
Description: | The thicker (1/2-inch) insulation was installed on the RF deck and the reference temperature was reduced from 50 C to 30 C to prevent thermal instability in winter. A change in the mixer temperature affects the instrument's calibration. A new calibration was automatically derived and applied after 1500 valid tip curves were acquired. During the period after the temperature was reset and before the calibration was updated, the previous calibration was used. |
Suggestions: | Reprocess the data with the new calibration. |
Measurements: | nsamwrlosC1.a1: more
nsamwrlosC1.b1: more |