DQRID : D010307.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/03/2000103002/15/20001735Incorrect
more
Subject:
SGP/EBBR/E8 - E8 EBBR AEM Stuck In One Position
DataStreams:sgp30ebbrE8.b1, sgp15ebbrE8.b1, sgp5ebbrE8.b1
Description:
When the AEM (automatic exchange mechanism) is stuck in one postion, 
half hour latent and sensible heat fluxes and the Bowen ratio
are incorrect.  Some of the raw EBBR measurements are also incorrect. 

I do not normally submit a DQR for this situation, as the qc checks for
AEM home signals normally indicate the problem.  However, the frequency
of the problem at this site seems to warrant DQRs.
Suggestions: 
Measurements:sgp5ebbrE8.b1:
  • temp_trh_top
  • temp_air_bottom
  • rh_bottom_fraction
  • temp_trh_bottom
  • vapor_pressure_bottom
  • rh_top_fraction
  • vapor_pressure_top
  • temp_air_top
more
sgp15ebbrE8.b1:
  • temp_air_right
  • temp_air_left
  • temp_trh_left_rr
  • rh_right_mv
  • rh_left_mv
  • temp_trh_right_rr
more
sgp30ebbrE8.b1:
  • vapor_pressure_bottom
  • latent_heat_flux
  • temp_air_bottom
  • temp_trh_bottom
  • rh_bottom_fraction
  • sensible_heat_flux
  • rh_top_fraction
  • vapor_pressure_top
  • bowen_ratio
  • temp_trh_top
  • temp_air_top
more

Close this window