DQRID : D030115.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/16/2002172012/10/20021720Does not affect quality
more
Subject:
SGP/EBBR/E12 - Reprocessed and Corrected: Net Rad Calibration Coefficients
DataStreams:sgp30ebbrE12.b1, sgp5ebbrE12.b1
Description:
DATA REPROCESSED and CORRECTED: 

Incorrect calibration used resulted in bad q values. Simple multipliers corrected the data 
and then sensible and latent heat fluxes were recalculated.

Positive net radiation measurements were multiplied by 0.972.
Negative net radiation measurements were multiplied by 1.0198.

The data were archived in April 2017. Please retrieve the reprocessed data for use in your 
analysis.

You may find these data using the ARM Discovery tool at the following link:
http://www.archive.arm.gov/discovery/#v/results/s/finst::ebbr/fsite::sgp/ffac::sgp.E12 


ORIGINAL DQR CATEGORY: Red( Incorrect)

REPROCESSED and CORRECTED DATA ARCHIVED:2017-04

REPROCESS IDENTIFICATION NUMBER: RID 228

---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------

The Net Radiometer coefficients were incorrect in the CR10
program when the replacement EBBR unit was installed on
08/16/02.

The data can be easily corrected; see the Suggestion area below.
Suggestions: 
Measurements:sgp5ebbrE12.b1:
  • net_radiation
more
sgp30ebbrE12.b1:
  • net_radiation
  • sensible_heat_flux
  • latent_heat_flux
more