DQRID : D030805.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/05/2003000008/06/20032359Does not affect quality
06/05/2003000006/11/20032359Does not affect quality
more
Subject:
TWP/AERI/C2 - Instrument maintenance, Sensor failure, hatch stuck
DataStreams:twpaerisummaryC2.a1
Description:
The rain sensors tell the hatch when to close so that the optical 
components are protected
                 from the weather.  When the rain sensors become either dirty or covered 
with salt spray, a 
                 hatch event is triggered. When dirty, they will give a low reading even 
when there is no rain.

                 Error flags also indicated the hatch was stuck in an intermediate 
position, neither open 
                 nor closed.  

                 The hatch closed under sunny conditions due to failure of the electronics 
of the hatch. No 
                 atmospheric spectral data are produced under this condition, which is 
equivalent to a rain 
                 condition.
Suggestions: 
The user should view the hatch condition to verify when the hatch was 

open, intermediate or 
                 closed and only use data when the hatch was open.
Measurements:twpaerisummaryC2.a1:
  • rainSensorIntensity
  • hatchOpen
more