Description: | The AEM home signals were too large at times during the indicated period;
it has been determined by the vendor that this is usually caused by electrical noise in
the vicinity. When the 30 minute data shows a home 30 signal of 35 or more, the bowen
ratio, and sensible and latent heat flux
measurements are incorrect
Removing the jumper between 4L and AG on the CR10 and adding a jumper to Multiplexer 2,
Set 6, between L1 and L2 fixed the problem. The
vendor intends to make this modification to all EBBR systems as they return
for re-calibration. |