Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/04/2003 | 0552 | 11/11/2003 | 1930 | Incorrect |
Subject: | SGP/MFRSR/E11 - Instrument failure |
DataStreams: | sgpmfrsrlangplotE11.c1, sgpmfrsrE11.b1, sgpmfrsrlangleyE11.c1 |
Description: | Beginning at 0552 UTC on 11/04, the MFRSR at E11 had frequent missing or erratic data. Corresponding to this, the head_temp readings and logger_volt readings were both flagged frequently (head_temp readings were erratic/noisy, and logger_volt readings were too high). Data quality was severely affected. Loggerboard replacement corrected the problem. |
Suggestions: | |
Measurements: | sgpmfrsrlangplotE11.c1: more
sgpmfrsrE11.b1: more sgpmfrsrlangleyE11.c1: more |