Subject:
NSA/MWR/C1  - no air temperature signal
DataStreams:nsamwrlosC1.b1, nsamwrC1.00, nsamwrtipC1.a1
Description:
When the new blower was upgraded by Radiometrics and reinstalled on the MWR, the air 
temperature sensor failed to properly report. It was determined that the wires carrying the 
signal to the analog board did not conform to the standard expected by the upgraded blower. 
The problem was corrected by changing the wiring and modifying the MWR software to read 
the signal from the appropriate corresponding channel.
Suggestions: 
Measurements:nsamwrtipC1.a1:
  • tkair
more
nsamwrC1.00:
  • Raw data stream - documentation not supported
more
nsamwrlosC1.b1:
  • tkair
more