Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/13/2003 | 1300 | 01/11/2005 | 1959 | Incorrect |
Subject: | SGP/MFRSR/E8 - Instrument problem, bad 415nm channel |
DataStreams: | sgpmfrsrlangplotE8.c1, sgpmfrsrlangleyE8.c1, sgpmfrsrE8.b1, sgpmfrsrE8.a0, sgpmfrsrE8.00 |
Description: | The measurements from the 415 nm channel are not correct. Data from this channel fluctuate, drop out and are often noisy. From the start of this DQR until around July 1, 2004, channel 415 data are consistently poor. Beginning in July we see 415 often looks OK. From this point until the end of the DQR channel 415 is sporadically bad, with longer periods of good appearing data as time moves on. |
Suggestions: | Data up to July 1, 2014 should not be used. After July 1, 2004 the data can be manually scanned to pick out good portions of the data. |
Measurements: | sgpmfrsrE8.b1: more
sgpmfrsrE8.a0: more sgpmfrsrE8.00: more sgpmfrsrlangleyE8.c1: more sgpmfrsrlangplotE8.c1: more |