Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/14/2004070602/08/20051836Incorrect
more
Subject:
SGP/SMOS/E8 - T/RH probe Intermittently failing
DataStreams:sgp1440smosE8.b1, sgp30smosE8.b1, sgp1smosE8.b1
Description:
Beginning of 12/14/2004 the RH% values at E8 became intermittently bad.  RH% values drop 
to 0% at times.  On 01/25/2005 @ 1820GMT a new probe was installed.  New probe RH% 
readings were over 104% in saturated conditions.  This probe was also replaced.
Suggestions: 
Use co-located EBBR T/RH data during times SMOS probe was failing.