Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/13/2005 | 1200 | 02/10/2005 | 1700 | Suspect |
Subject: | SGP/SIRS/E6 - Questionable data due to Ice Storm damage |
DataStreams: | sgpsirsE6.b1, sgpsirsE6.00, sgpsirs20sE6.a0 |
Description: | Several SIRS readings at E6 readings have appeared questionable since the ice storm in early January. Starting on 01/13 maintenance found the E6 tracker inoperative. It was reset and put back into serivice. Shortly after, on 01/16 data from the DSH instrument (PSP) became questionable possibly due to obstructed or failed ventilator fan. On 02/10 1645 GMT maintenance found the DSH instrument ventilator fan failed - the fan was replaced, the E6 tracker was also realigned. Due to early january ice storms, the tracker alignment issues and a failed ventilator fan - All tracker mounted and global DS instrument measurements are suspect for this time frame. |
Suggestions: | Disregard this data or pre-review data before use. |
Measurements: | sgpsirsE6.b1: more
sgpsirs20sE6.a0: more sgpsirsE6.00: more |