DQRID : D050822.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/05/2005200408/07/20050555Does not affect quality
more
Subject:
SGP/MWR/B1 - Reprocessed and Corrected: mixer temp adjustment
DataStreams:sgpmwrlosB1.b1
Description:
DATA REPROCESSED and CORRECTED: 

The RF deck temperature was increased from 50 to 52 C to prevent thermal instability in 
summer. This resulted in a change in the calibration. The data were reprocessed with the 
calibration coefficients that were automatically derived after the reference temperature 
tkmx was adjusted. 

ORIGINAL DQR CATEGORY: Yellow( Suspect)

REPROCESSED and CORRECTED DATA ARCHIVED:2016-03-26

REPROCESS IDENTIFICATION NUMBER: RID 207
 
---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------
The RF deck temperature was increased from 50 to 52 C to prevent thermal instability in 
summer.  This resulted in a change in the calibration.  The data need to be reprocessed 
with the calibration coefficients that were automatically derived after the reference 
temperature, tkmx, was adjusted.
Suggestions: 
Measurements:sgpmwrlosB1.b1:
  • tc31
  • tbsky23
  • tnd23
  • tbsky31
  • tnd_nom31
  • tnd31
  • vap
  • tc23
  • liq
  • tnd_nom23
more