Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/05/2005 | 1842 | 04/19/2006 | 1713 | Suspect |
Subject: | SGP/CEIL/C1 - poor signal |
DataStreams: | sgpceilC1.b1 |
Description: | When parameters were changed (set noise_h2_compensation to "ON" and units to "METERS") to correct a low amplitude problem, the backscatter plots began displaying horizontal stripes indicating detector "ringing". However, the cloud base heights are not affected. The problem was corrected when the instrument was replaced. |
Suggestions: | |
Measurements: | sgpceilC1.b1: more
|