Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/16/2006 | 0300 | 06/20/2006 | 2100 | Incorrect |
Subject: | TWP/MMCR/C1 - Reprocess: Reflectivities too high |
DataStreams: | twpmmcrmomC1.b1 |
Description: | After the upgrade to PIRAQ-III, the reflectivies are too high by ~ 23 dBZ. The problem was caused by a failure of one of the power supplies that turned off one of the rf switches. The data can be used but 23 dB should be subtracted from the reported values. Due to the unknown exact isolation the rf switch offers, one should assume that the reflectivity will then be accurate to +/- 3 dBZ. |
Suggestions: | |
Measurements: | twpmmcrmomC1.b1: more
|