Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/16/2006 | 0000 | 07/22/2006 | 0600 | Does not affect quality |
Subject: | SGP/MWR/B1 - Reprocessed: Corrected calibration |
DataStreams: | sgpmwrlosB1.b1, sgpmwrB1.00, sgpmwrtipB1.a1 |
Description: | The radiometer was thermally unstable. Calibration coefficients for the time interval were re-derived after screening for bad values of TKxc and the data were reprocessed. Reprocessed MWRLOS data were archived January 2007. MWRTIP data will not be reprocessed. |
Suggestions: | Retrieve the reprocessed data for use in your analysis. |
Measurements: | sgpmwrB1.00: more
sgpmwrlosB1.b1: more sgpmwrtipB1.a1: more |