DQRID : D080115.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/21/2007000012/23/20070000Incorrect
more
Subject:
NSA/MWRP/C1 - Radiometer thermally unstable
DataStreams:nsamwrpC1.b1
Description:
Between Dec 21 and Dec 23 the external temperature dropped to - 38 C. This caused a 
problem with the stability of the noise diode mount temperature and consequent problem with 
calibration.
Suggestions: 
Avoid data between 12/21 and 12/23.
Measurements:nsamwrpC1.b1:
  • relativeHumidity
  • liquidWaterContent
  • waterVaporMixingRatio
  • virtualTemperature
  • brightnessTemperature
  • dewpointTemperature
  • totalPrecipitableWater
  • totalPrecipitableWater2
  • waterVaporDensity
  • liquidWaterPath
  • temperature
  • liquidWaterPath2
more