DQRID : D080402.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/01/2007000010/13/20071615Suspect
more
Subject:
FKB/MWRHF/M1 - Questionable calibration
DataStreams:fkbmwrhfM1.b1
Description:
The instrument completed a tip curve calibration on June 30 and it did not calibrate again 
until Oct 13. Data during this time display large residuals (~ 6 K) when compared with 
model computations. It is likely that the tip curve calibration, although successful, 
wasn't good enough to have good accuracy.

Since this instrument is still under testing the users should use the data with caution. 
Although the accuracy of the model itself is still debated, large residuals between the 
measurements and the model between June and October are probably in large part due to the 
poor calibration.
Suggestions: 
Measurements:fkbmwrhfM1.b1:
  • tbsky150
  • tbsky90
more