Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/08/2007 | 2100 | 05/25/2007 | 0000 | Suspect |
Subject: | SGP/QCRAD/E24 - Intermittently noisy Tc/Td |
DataStreams: | sgpqcrad1longE24.s1, sgpqcrad1longE24.s2, sgpqcrad1longE24.c1, sgpqcrad1longE24.c2 |
Description: | Case and dome temperatures for Longwave fluxes are intermittently noisy during this period corresponding to additional noise in the longwave fluxes. No instrument DQRs exist documenting this problem. |
Suggestions: | Longwave fluxes have higher uncertainties during this time period, use with caution. |
Measurements: | sgpqcrad1longE24.s1: more
sgpqcrad1longE24.c1: more sgpqcrad1longE24.c2: more sgpqcrad1longE24.s2: more |