Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/04/2003 | 0000 | 02/06/2007 | 2359 | Suspect |
Subject: | SGP/MET/E9 - Temperature occasionally failed 6-month field calibration checks |
DataStreams: | sgpmetE9.b1 |
Description: | The temperature/humidity sensor was not calibrated annually during this time. The temperature reading failed the 6-month field calibration check four times during this period. Upon removal, the field calibration check reported the temperature sensor may have been reading high (< 1 deg C). In general, the relative humidity may have been subject to drift during this time. The manufacturer characterizes sensor drift at <1% / year. |
Suggestions: | Please contact the mentor for specific information regarding the field calibration check results, if necessary. |
Measurements: | sgpmetE9.b1: more
|