Subject:
SGP/MET/E21 - Temperature/humidity sensor occasionally failed 6-month field calibration 
checks
DataStreams:sgpmetE21.b1
Description:
The temperature/humidity sensor has not been calibrated since 2/27/2007. The temperature 
reading failed four 6-month field calibration checks during this time, but has passed 
field calibration checks after 10/31/2013. 

In general, the relative humidity may have been subject to drift during this time. The 
manufacturer characterizes sensor drift at <1% / year.
Suggestions: 
Please contact the mentor for specific information regarding the field calibration 

check results, if necessary.