DQRID : D080211.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/31/2008000002/08/20080248Suspect
more
Subject:
TWP/SKYRAD/C1 - Failed PSP ventilator fan
DataStreams:twpskyrad60sC1.b1, twpskyrad20sC1.a0
Description:
On 1/31 the ventilator for the PSP measuring downwelling_shortwave_hemispheric irradiance 
failed.  The DQ effect of this failure on the measurement of down_short_hemisp is not 
possible to fully quantify but it has been noted that during these failures the PSP 
nighttime values can be several (1-3W/m2) smaller than usual.  Another effect is slower thermal 
response (1hr) of the PSP instrument due to daytime heating and cooling of the ventilator 
housing.
Suggestions: 
Use data with caution.
Measurements:twpskyrad60sC1.b1:
  • down_short_hemisp_std
  • down_short_hemisp_min
  • down_short_hemisp
  • down_short_hemisp_max
more
twpskyrad20sC1.a0:
  • inst_global
more