DQRID : D090213.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/09/2008050012/28/20082359Does not affect quality
more
Subject:
HFE/MWRHF/M1 - Calibration updates
DataStreams:hfemwrhfM1.b1
Description:
The MWRHF calibration algorithm updates the calibration parameters every time a new 
successful tip curve is acquired. 

This results in small jumps (1-2 K) in the brightness temperatures each time calibration 
is updated. During times of frequent calibrations the data appear discontinuous with short 
data segments interrupted by small jumps.
Suggestions: 
Measurements:hfemwrhfM1.b1:
  • tbsky150
  • tbsky90
more