Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/21/2008 | 1300 | 11/10/2008 | 2359 | Suspect |
Subject: | SGP/QCRAD/E11 - Instrument multiplexer failure |
DataStreams: | sgpqcrad1longE11.c1, sgpqcrad1longE11.c2, sgpqcrad1longE11.s2, sgpqcrad1longE11.s1 |
Description: | The multiplexer failed intermittently during this time. The down_short_hemisp, up_long_hemisp and down_long_hemisp were very noisy. The SIRS multiplexer wiring or mux itself was failing during this time. Ops replaced the mux on 12/2. The problem is documented in instrument DQR D081029.2 |
Suggestions: | Use caution with SWup, SWdn, LWdn flux data during this period. |
Measurements: | sgpqcrad1longE11.c1: more
sgpqcrad1longE11.s1: more sgpqcrad1longE11.c2: more sgpqcrad1longE11.s2: more |