DQRID : D100608.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/01/2009000011/28/20102330Suspect
more
Subject:
TWP/CSPHOT/C2 - 340 nm filter degradation.
DataStreams:twpcsphotaotfiltC2.a1, twpcsphotalm1duboC2.a1, twpcsphotaotC2.a1, twpcsphotaotfiltqaC2.a1
Description:
There was 340 nm filter degradation in the Nauru Cimel Sunphotometer starting 11/1/2009.  
The 340 AOD was in disagreement with other channels sporadically starting 11/1/2009 and 
consistently off from the other channels by March 2010.  Therefore 340 AOD channel should 
not be used until the unit is replaced.  

The 340 nm filter will be changed when the instrument is sent back for calibration.
Suggestions: 
Measurements:twpcsphotaotfiltqaC2.a1:
  • aot_340
  • Angstrom340_440
  • TripletVar_340
more
twpcsphotaotC2.a1:
  • aot_340
  • TripletVar_340
  • Angstrom340_440
more
twpcsphotalm1duboC2.a1:
  • AOT_340
more
twpcsphotaotfiltC2.a1:
  • Angstrom340_440
  • aot_340
  • TripletVar_340
more

Close this window