Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
06/21/2010 | 1930 | 07/26/2010 | 1900 | Incorrect |
Subject: | SGP/MFRSR/E11 - Shadowband Misalignment |
DataStreams: | sgpmfrsrlangplotE11.c1, sgpmfrsrE11.b1, sgpmfrsrlangleyE11.c1, sgpmfrsrE11.00, sgpmfrsraod1michE11.c1, sgpmfrsrE11.a0 |
Description: | The instrument mounting plate became loose causing the instrument to go out of alignment. Maintenance resolved the problem on 20100726 just after 1900 UTC. Affected hours are from 1930 to sunset. |
Suggestions: | |
Measurements: | sgpmfrsrlangplotE11.c1:
sgpmfrsrE11.00:
sgpmfrsrlangleyE11.c1:
sgpmfrsrE11.a0:
sgpmfrsrE11.b1:
sgpmfrsraod1michE11.c1:
|