DQRID : D101229.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/26/2010030011/26/20101000Suspect
12/02/2010020012/03/20102359Suspect
11/14/2010070011/15/20100400Suspect
12/11/2010030012/11/20101400Suspect
12/25/2010000012/26/20102030Suspect
12/23/2010000012/23/20101500Suspect
11/11/2010000011/11/20100800Suspect
more
Subject:
NSA/SKYRAD/C2 - Suspect PIR1
DataStreams:nsaskyrad60sC2.b1, nsaskyrad20sC2.a0
Description:
PIR1 (down_long_hemisp) data are suspect during these time ranges due to the large 
differences between PIR1 and PIR2 (i.e. greater than 10W/m2).  Maintenance on 12/26 indicated a 
failed ventilator heater connector on PIR1.  After 12/26 PIR1 and PIR2 returned to 
tracking each other within 10W/m2 as typical.
Suggestions: 
Use PIR2 (down_long_hemisp2) data. The QC flags appear to effectively call out the 

bad data periods.
Measurements:nsaskyrad20sC2.a0:
  • inst_down_long_hemisp1_case_resist
  • inst_down_long_hemisp1_tp
  • inst_down_long_hemisp1_dome_resist
more
nsaskyrad60sC2.b1:
  • down_long_hemisp1_max
  • down_long_hemisp1_std
  • inst_down_long_hemisp1_tp
  • down_long_hemisp1_min
  • inst_down_long_hemisp1_dome_temp
  • down_long_hemisp1
  • inst_down_long_hemisp1_case_temp
more