DQRID : D120510.26
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/19/1999220005/27/19992000Suspect
01/01/1999000003/18/19991500Suspect
06/09/2010090006/12/20101700Suspect
08/17/2010060008/24/20100000Suspect
more
Subject:
SGP/QCRAD/E6  - Tc/Td noisy
DataStreams:sgpqcrad1longE6.c1, sgpqcrad1longE6.s1, sgpqcrad1longE6.s2, sgpqcrad1longE6.c2
Description:
Case and dome temperatures for longwave fluxes are noisy during this period corresponding 
to additional noise in the longwave fluxes.

No instrument DQRs exist documenting this problem.
Suggestions: 
Longwave fluxes have higher uncertainties during this time period, use with caution.
Measurements:sgpqcrad1longE6.c1:
  • LWdnTd
  • down_long_hemisp
  • up_long_hemisp
  • LWdnTc
  • LWupTd
  • LWupTc
more
sgpqcrad1longE6.s1:
  • LWdnTd
  • LWdnTc
  • LWupTc
  • LWupTd
  • up_long_hemisp
  • down_long_hemisp
more
sgpqcrad1longE6.c2:
  • LWdnTd
  • LWdnTc
  • LWupTd
  • up_long_hemisp
  • LWupTc
  • down_long_hemisp
more
sgpqcrad1longE6.s2:
  • up_long_hemisp
  • down_long_hemisp
  • LWdnTc
  • LWdnTd
  • LWupTd
  • LWupTc
more

Close this window