Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/11/2011 | 0000 | 04/22/2011 | 2359 | Incorrect |
Subject: | SBS/MWR3C/M1 - Radiometer unstable |
DataStreams: | sbsmwr3cM1.b1 |
Description: | The noise diode was unstable and the instrument was unable to calibrate. In addition the instrument did not have a rain mitigation system during the SBS deployment, therefore snow and dew stayed on the lens affecting the measurements. |
Suggestions: | |
Measurements: | sbsmwr3cM1.b1: more
|