Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/12/2009000005/11/20111520Incorrect
more
Subject:
SGP/IRT/E12 - Reprocess: Calibration error
DataStreams:sgpirtE12.b1
Description:
The measured sky temperature was lower than other SGP facilities because the original 
calibration, as set be the vendor, was wrong. The IRT was replaced and returned to the vendor.
Suggestions: 
These data will be reprocessed after the calibration error is determined
Measurements:sgpirtE12.b1:
  • sky_ir_temp_min
  • sky_ir_temp
  • sky_ir_temp_max
more