DQRID : D130501.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/11/2012075002/08/20131957Does not affect quality
more
Subject:
SGP/IRT/C1 - 25m IRT ground surface temperature
DataStreams:sgpirt25mC1.b1
Description:
The 10m IRT was measuring lower ground surface temperature than the 25m IRT. The 25m lens 
was cleaned on 12/24/2012 and replaced on 2/8/2013 but the difference in temperature 
appears to be due to the seasonal difference in surface cover - tall grass vesus bare soil.
Suggestions: 
Measurements:sgpirt25mC1.b1:
  • sfc_ir_temp_max
  • sfc_ir_temp
  • sfc_ir_temp_min
  • sfc_ir_temp_std
more