Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/24/2012 | 0000 | 04/24/2013 | 0000 | Does not affect quality |
Subject: | SGP/AOS/C1 - Reprocessed: Bad PSAP flow cal |
DataStreams: | sgpnoaaaosC1.b1 |
Description: | There was a leak in the flow calibration device that resulted in a lower than expected flow calibration. The data were reprocessed with a flow calibration that is the average of the flow cals taken before and after this time period. |
Suggestions: | Reprocessed November 2013 |
Measurements: | sgpnoaaaosC1.b1: more
|