Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/16/2013 | 1631 | 12/18/2013 | 1938 | Does not affect quality |
Subject: | SGP/SIRS/E37 - Reprocessed: Incorrect LW calibration coefficients |
DataStreams: | sgpsirsE37.b1 |
Description: | During the time range specified the UIR and DIR serial numbers were correct but incorrect calibration coefficients were applied in the datalogger. up_long_hemisp (SN# 30834F3) wrong k1 cal :0.26431 W/(m^2*uV) correct k1 cal: 0.2667 W/(m^2*uV) down_long_hemisp (SN# 30835F3) wrong k1 cal :0.2384 W/(m^2*uV) correct k1 cal: 0.2457 W/(m^2*uV) |
Suggestions: | Data were reprocessed in February 2014. |