DQRID : D141022.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/12/2012000005/16/20130004Does not affect quality
more
Subject:
MAG/PSAP/M1 - Impactor not operational
DataStreams:magaospsap3wM1.b1
Description:
All data in this period should be considered to have a 10-um cutoff.  This was confirmed 
by observation of the PSAP flow.
Suggestions: 
Use data for 10um analysis only
Measurements:magaospsap3wM1.b1:
  • qc_mass_flow
  • qc_Ba_R_PSAP3W
  • qc_transmittance_R
  • mass_flow
  • dilution_flow_valve_position
  • impactor_setting
  • time
  • dilution_flow_setpoint
  • dilution_flow
  • base_time
  • transmittance_R
  • qc_Ba_B_PSAP3W
  • alt
  • sample_length
  • transmittance_G
  • qc_impactor_setting
  • time_offset
  • lat
  • Ba_R_PSAP3W
  • transmittance_B
  • qc_transmittance_G
  • qc_Ba_G_PSAP3W
  • Ba_G_PSAP3W
  • lon
  • qc_transmittance_B
  • Ba_B_PSAP3W
  • qc_sample_length
more

Close this window