DQRID : D140103.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/27/2013164801/02/20141923Does not affect quality
more
Subject:
MAO/SKYRAD/M1 - Reprocessed: IRT configuration error
DataStreams:maoskyrad60sM1.b1
Description:
SKYRAD IRT sky temperature measurements were ~40K too low. The analog output configuration 
of the IRT was automatically set to incorrect values by the serial collection script. An 
error in the name of the startup file was corrected.

The data were reprocessed to correct the configuration/calibration and the reprocessed 
data were archived in February, 2014.
Suggestions: 
Measurements:maoskyrad60sM1.b1:
  • sky_ir_temp_max
  • sky_ir_temp_min
  • sky_ir_temp
more