DQRID : D140211.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/23/2013160001/07/20141734Incorrect
more
Subject:
SGP/SIRS/E11 - Ventilator Fan Failure
DataStreams:sgpsirs20sE11.a0, sgpsirsE11.b1
Description:
Beginning on 12/23/2013 approx. 16:00GMT the down_short_hemisp (PSP instrument) ventilator 
fan failed. Site Ops repaired the ventilator and returned the instrument to service on 
1/7/2014.

The down_short_hemisp data are not recoverable during this time range.  However, users can 
reconstruct the data using the following relationship between the down_short_hemisp and 
the short_direct_normal and short_diffuse_hemisp data for each one minute measurement:

down_short_hemisp = short_direct_normal * cos(zenith angle) + short_diffuse_hemisp

Problem should not adversely affect any overlying VAPs such as the QCRAD that will use 
this relationship as determined by QC flags.
Suggestions: 
Measurements:sgpsirs20sE11.a0:
  • inst_global
more
sgpsirsE11.b1:
  • down_short_hemisp
  • down_short_hemisp_std
  • down_short_hemisp_min
  • down_short_hemisp_max
more