Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/24/2013 | 0000 | 03/24/2014 | 1545 | Does not affect quality |
Subject: | SGP/SIRS/C1 - Sensor alignment/level |
DataStreams: | sgpsirsC1.a2, sgpsirs60sC1.a1, sgpsirsC1.a0, sgpsirsC1.a1, sgpsirsC1.00, sgpsirsC1.b1, sgpsirsC1.01, sgpsirs20sC1.a0 |
Description: | During the time range specified the downwelling shortwave DS radiometer (down_short_hemisp measurement) was slightly out of level. The affects on data quality are within the uncertainly of the instrument. However, the derived DS value (component sum of direct and diffuse shortwave) should be used during this time period. Any of the associated VAPs (QCRAD, Central Facility Best Estimate) use the component sum by default and will not be adversely affected. |
Suggestions: | The component sum of direct and diffuse shortwave should be used during this time period. |
Measurements: | sgpsirs60sC1.a1: more
sgpsirsC1.a2: more sgpsirsC1.01: more sgpsirsC1.b1: more sgpsirsC1.a0: more sgpsirsC1.a1: more sgpsirsC1.00: more sgpsirs20sC1.a0: more |