DQRID : D150109.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/01/2014030012/30/20141705Suspect
more
Subject:
SGP/IRT/E12 - Instrument contamination
DataStreams:sgpirtE12.b1
Description:
The sky temperature had an apparent bias of ~15 K higher than other sites. The lens and 
mirror were cleaned.
Suggestions: 
Measurements:sgpirtE12.b1:
  • sky_ir_temp_min
  • sky_ir_temp
  • sky_ir_temp_max
more