Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
10/18/2013 | 0000 | 07/10/2014 | 1010 | Incorrect |
Subject: | ENA/ACSM/C1 - Bad ACSM data |
DataStreams: | enaaosacsmC1.a1 |
Description: | After initial startup of the container at ENA, the ACSM was behaving erratically. It was determined that the filter switching valve had failed. Because of the difficulties in getting parts to the site, the replacement valve was not delivered until June, 2014. After delivery, there were tuning issues with the instrument. Instrument tuning was accomplished remotely on 2014-07-10 and resulted in good data. |
Suggestions: | Do not use data |
Measurements: | enaaosacsmC1.a1: more
|