Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/19/2015 | 0419 | 03/02/2015 | 2121 | Incorrect |
04/06/2015 | 0711 | 04/11/2015 | 1833 | Incorrect |
Subject: | MAO/AOS/M1 - Weak solder connection in dilution flow meter comport cable |
DataStreams: | maoaosccn100M1.a1, maoaosclap3wM1.a1, maoaoscpcM1.a1 |
Description: | The dilution flow meter reading was erratic. Plots of the data suggest that the flow varied by about 5%. Problem was corrected when wire was re-soldered. |
Suggestions: | For more precise data use the b1 level data as correction has been applied to this data. For a good approximation multiply concentrations, scattering and absorption coefficients by 1.35 to obtain approximate dilution correction. |
Measurements: | maoaoscpcM1.a1: more
maoaosclap3wM1.a1: more maoaosccn100M1.a1: more |