Subject:
MAO/AOS/M1 - Weak solder connection in dilution flow meter comport cable
DataStreams:maoaosccn100M1.a1, maoaosclap3wM1.a1, maoaoscpcM1.a1
Description:
The dilution flow meter reading was erratic. Plots of the data suggest that the flow 
varied by about 5%. Problem was corrected when wire was re-soldered.
Suggestions: 
For more precise data use the b1 level data as correction has been applied to this 

data. For a good approximation multiply concentrations, scattering and absorption 
coefficients by 1.35 to obtain approximate dilution correction.