Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
03/24/2015 | 1600 | 09/02/2015 | 1941 | Missing |
Subject: | MAO/CLAP/M1 - Instrument removed for repair |
DataStreams: | maoaosclap3wM1.a1 |
Description: | A problem with the internal instrument flow meter was discovered as the flow calibration had shifted substantially from the previous calibration. |
Suggestions: | Use PSAP data |
Measurements: | maoaosclap3wM1.a1: more
|