DQRID : D150415.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/24/2015160009/02/20151941Missing
more
Subject:
MAO/CLAP/M1 - Instrument removed for repair
DataStreams:maoaosclap3wM1.a1
Description:
A problem with the internal instrument flow meter was discovered as the flow calibration 
had shifted substantially from the previous calibration.
Suggestions: 
Use PSAP data
Measurements:maoaosclap3wM1.a1:
  • transmittance_R
  • alt
  • impactor_setting
  • base_time
  • clap_flow_rate
  • lon
  • time_offset
  • Ba_G_CLAP3W
  • transmittance_G
  • Ba_B_CLAP3W
  • sample_length
  • transmittance_B
  • filter_id
  • lat
  • time_bounds
  • Ba_R_CLAP3W
  • instrument_flags
  • time
  • active_filter_number
more

Close this window