DQRID : D150420.9
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/21/2015100303/21/20151143Incorrect
03/07/2015234003/08/20150141Incorrect
03/14/2015065303/14/20151221Incorrect
03/31/2015160203/31/20151705Incorrect
03/07/2015145403/07/20151623Incorrect
03/23/2015223803/23/20152354Incorrect
03/08/2015024903/08/20150417Incorrect
03/27/2015104903/27/20151305Incorrect
03/31/2015182503/31/20152357Incorrect
03/23/2015052303/23/20151210Incorrect
03/30/2015040403/30/20150544Incorrect
03/13/2015151003/13/20151649Incorrect
03/21/2015010303/21/20150538Incorrect
03/13/2015095803/13/20151149Incorrect
03/18/2015185503/18/20152332Incorrect
03/06/2015152403/06/20151711Incorrect
more
Subject:
MAO/PSAP/S1 - Gaps greater than 1 hour due to transmission
DataStreams:maoaospsap3wS1.b1, maoaospsap3wS1.a1
Description:
Manufacturer recommends information valid only for transmission greater than 0.7.  Other 
periods are oscillations after filter changes and pump off.
Suggestions: 
Do not use data
Measurements:maoaospsap3wS1.a1:
  • impactor_setting
  • filter_id
  • Ba_R_PSAP3W
  • Ba_B_PSAP3W
  • transmittance_R
  • Ba_G_PSAP3W
  • psap_flow_rate
  • transmittance_G
  • transmittance_B
  • sample_length
more
maoaospsap3wS1.b1:
  • TrG
  • Ba_R_PSAP3W
  • Ba_G_PSAP3W
  • TrB
  • Ba_B_PSAP3W
  • TrR
more