DQRID : D150612.10
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/22/2015000006/10/20152359Suspect
more
Subject:
ENA/CSPHOT/C1 - Obstruction in collimator tubes
DataStreams:enacsphotaotfiltC1.a1, enacsphotalmC1.a1, enacsphotppC1.a1, enacsphotaotC1.a1
Description:
Data are questionable due to possible obstruction in collimator tubes.
Suggestions: 
During this time period mentor recommends using the quality assured level 2.0 AOT 

data from AERONET when available. Level 2.0 data and post-calibration level 1.5 data should 
become available in 2016. Can also use MFRSR data if available.
Measurements:enacsphotppC1.a1:
  • pp_rad
more
enacsphotaotfiltC1.a1:
  • TripletVar_440
  • aot_1020
  • Angstrom380_500
  • Angstrom500_870
  • aot_380
  • aot_340
  • Angstrom340_440
  • Angstrom440_675
  • TripletVar_870
  • TripletVar_670
  • Angstrom440_870
  • TripletVar_1640
  • aot_870
  • TripletVar_380
  • TripletVar_340
  • TripletVar_500
  • Angstrom440_675_Polar
  • aot_1640
  • pwv
  • aot_440
  • aot_500
  • WaterError
  • TripletVar_1020
  • aot_670
more
enacsphotalmC1.a1:
  • alm_rad
more
enacsphotaotC1.a1:
  • Angstrom500_870
  • Angstrom380_500
  • TripletVar_380
  • TripletVar_340
  • Angstrom340_440
  • aot_380
  • TripletVar_500
  • aot_340
  • TripletVar_1020
  • aot_500
  • TripletVar_1640
  • aot_1640
  • Angstrom440_675
  • Angstrom440_870
  • Angstrom440_675_Polar
  • WaterError
  • TripletVar_440
  • aot_440
  • TripletVar_670
  • TripletVar_870
  • aot_1020
  • aot_670
  • aot_870
more

Close this window